Skip to main content

Table 1 Characteristics of tested devices, baseline characteristics of patients, and differences in usability

From: Dry-powder inhalers in patients with persistent airflow limitation: usability and preference

 

Breezhaler

Genuair

Handihaler

p

Device characteristics

    

Manoeuvres (n)a

7

3

4

 

 Time spent by nurse with patients during trainingb (min)

1st attempt

2.8 ± 0.1

1.3 ± 0.1

1.4 ± 0.1

p < 0.0001

2nd attempt

4.2 ± 0.2

1.8 ± 0.1

2 ± 0.2

p < 0.0001

3rd attempt

5.0 ± 0.3

2.1 ± 0.2

2.7 ± 0.2

p < 0.0001

Patient characteristics

     

 Patients (N)

333

237

223

 

 Age (years)

55.2 ± 18.3

54.8 ± 17.5

57.8 ± 17.9

NS

 Sex (% male)

46.5 %

49.4 %

43.9 %

NS

 Disease (% COPD)

47.4 %

47.7 %

51.6 %

NS

 Previous experience with DPI

63.7 %

66.7 %

64.6 %

NS

 Previous instruction to use of DPI

60.7 %

64.1 %

62.8 %

NS

Usability

    

 N. attempts before achieving the first proper inhalation

2.6 ± 1.1

1.6 ± 0.8

2.5 ± 1.1

p < 0.0001

 Successful rate at the first attempt (%)

18.0 %

55.7 %

19.3 %

p < 0.0001

 Total time required for the first proper inhalationb (min)

10.3 ± 5.0

2.5 ± 1.6

4.9 ± 2.7

p < 0.0001

  1. DPI dry powder inhaler
  2. aNumber of operations to prepare the inhalation
  3. bIncluding patient time and nurse demonstration